Overview of Human ET-1 (Endothelin1) ELISA Kit for Cometic & Personal Care Research 

Product name: Human ET-1(Endothelin1) ELISA Kit
Reactivity: Human
Alternative Names: ET1; PPET1; Preproendothelin-1; Big endothelin-1; EDN1
Assay Type: Sandwich
Sensitivity: 2.71 pg/mL
Standard: 500 pg/mL
Range: 7.82-500 pg/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Endocrinology;Cardiovascular biology;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Endothelin1(ET-1). Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Endothelin1(ET-1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Endothelin1(ET-1), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Endothelin1(ET-1) in the samples is then determined by comparing the OD of the samples to the standard curve.

[accordions]
[accordion title= “Standard Curve“]

Concentration (ng/mL) OD Corrected OD
500.00 2.057 1.951
250.00 1.498 1.392
125.00 0.983 0.877
62.50 0.755 0.649
31.25 0.507 0.401
15.63 0.329 0.223
7.82 0.208 0.102
0.00 0.106 0.000

standard curve of Human ET-1(Endothelin1) ELISA Kit

[/accordion]

[accordion title= “Precision“]

Intra-assay Precision (Precision within an assay):CV%<8% Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays):CV%<10% Three samples of known concentration were tested in forty separate assays to assess inter-assay precision. [/accordion] [accordion title= "Recovery“]

Matrices listed below were spiked with certain level of recombinant ET-1 and the recovery rates were calculated by comparing the measured value to the expected amount of ET-1 in samples.

 

Matrix Recovery range Average
serum(n=5) 86-99% 92%
EDTA plasma(n=5) 80-95% 87%
Heparin plasma(n=5) 90-105% 97%

[/accordion]

[accordion title=”Linearity“]

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ET-1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

 

Matrix 1:2 1:4 1:8 1:16
serum(n=5) 87-91% 85-93% 89-97% 85-93%
EDTA plasma(n=5) 89-97% 85-94% 87-96% 85-97%
Heparin plasma(n=5) 85-94% 87-96% 92-101% 82-98%